10. RAD-IO2 CALIBRATION
Enable Calibration:
Goto Settings Menu

Change calibration option to on

Calibration page should now be available

Click Read to get current calibration constants

Equipment needed for calibration:
TC and AIN:
Fluke 7562A or equivalent
AOUT
6 ½ digit multimeter such as Keysight 34465A or Keithley DMM6500 or equivalent

Read: Reads current calibration data
Store: Writes displayed calibration data to device
Export: Saves copy of current calibration data
Import: Loads saved calibration data
Clear: Clears existing calibration data, Will reset so all calibrations points have zero error.
Manual: Requires the user to enter the calibration value
Interactive: Reads actively reads input. Pressing enter or tab will advance to the next bank then next point. This is the fastest way to calibrate a device.